This spring-loaded thermocouple is designed for threaded blind-hole meaurements, surfaces subject to vibration/oscillation, and any application where positive contact for good measurement is required. The sensing tip of this device consists of a thin ribbon thermocouple element overlapped onto a phenolic rod. This rod is spring-loaded within the threaded stainless steel thermowell body. Thus, the sensing tip can be depressed up to an 1/8 inch while it continuously maintains good thermal contact to the object whose temperature is being measured. The combination of the thin ribbon thermocouple and low thermal conductivity of the phenloc rod yield response times in the low 200 milliseconds. This combination also provides temperature measurements without conduction errors. The true temperature of the test item is monitored.
All units come with a Delrin locking element (slug), which when used resists loosening due to vibration. However, this limits the upper temperature limit to 300 degrees Fahrenheit. The Delrin slug can easily be removed with a pair of plyers, increasing the upper temperature limit to 450 degrees Fahrenheit. There are several other techniques one can use to resist vibration. Contact an engineer to discuss these options. Also, if higher temperatures are required please provide us with your particular requirement.
|